Specifications

Dimensions: 

3.2” W x 5.1” H x 1.3” D

Weight:

8 oz.

Pressure Options: 

-40 to -125 mmHg or -.77 to -2.4 PSI
-5.3 to -16.6 kPa

Therapy Delivery Modes:

Continuous & Intermittent

Battery Type:

Alkaline

Battery Life: 

3 weeks or more

Patient Enclosure

 

Leakage Current: 

N/A (Battery powered device)

Storage Conditions:

Temperature Range: -20 °C to +60 °C

Relative Humidity Range: 0 to 95% non-condensing

Operating Conditions:

Temperature Range: +5 °C to +40 °C

Relative Humidity Range: 0 to 95% non-condensing

Altitude Range

Operating: -1000 ft to +10,000 ft

Storage:-1000 ft to +18,000 ft

IEC Classification

- Medical Equipment
- Equipment not suitable for use in the presence of flammable anesthetic mixture
- Type B, Applied Part
- Class II
- IPX0

 CAUTION:  Federal (USA) law restricts these devices to sale by or on the order of a physician

ELECTROMAGNETIC COMPATIBILITY

Electromagnetic Compatibility- Medical electrical equipment needs special precautions regarding electromagnetic compatibility (EMC) and needs to be installed and put into service according to the EMC information provided in this user manual. Portable and mobile radio frequency (RF) communications equipment can affect medical electrical equipment.

 

Guidance and Manufacturer’s Declaration - Electromagnetic Emissions

The NPD 1000 is intended for use in the electromagnetic environments specified below. The customer or the user of the NPD 1000 should assure it is used in such an environment.

 

Emissions Test

Compliance

Guidance

RF Emissions CISPR 11

Group 1

The NPD 1000 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

RF Emissions CISPR 11

Class B

The NPD 1000 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

Harmonic Emissions  IEC 61000-3-2

Not Applicable

 

Voltage Fluctuations/ Flicker Emissions IEC 61000-3-3

Not Applicable

 

 

 

 

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

The NPD 1000 is intended for use in the electromagnetic environments specified below. The customer or the user of the NPD 1000 should assure it is used in such an environment.

Immunity Test

IEC 60601
Test Level

Compliance Level

Guidance

Electrostatic Discharge 
IEC 61000- 4-2

± 6 kV Contact 
± 8 kV Air

± 6 kV Contact  ± 8 kV Air

Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%

Electrical Fast Transient/
Burst  IEC 61000-4-4

Not Applicable

Not Applicable

The NPD 1000 has no connection to Mains

Surge IEC 61000-4-5

Not Applicable

Not Applicable

The NPD 1000 has no connection to Mains

Voltage Dips, Short Interrupts, & Variations on Power Supply Lines                IEC 61000-4-11

Not Applicable

Not Applicable

The NPD 1000 has no connection to Mains

Power Frequency Magnetic Fields IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment

 

ELECTROMAGNETIC COMPATIBILITY CON’T

 

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

The NPD 1000 is intended for use in the electromagnetic environments specified below. The customer or the user of the NPD 1000 should assure it is used in such an environment.

Immunity Test

IEC 60601
Test Level

Compliance Level

Guidance