Dimensions:
|
3.2” W x 5.1” H x 1.3” D
|
|
Weight:
|
8 oz.
|
|
Pressure Options:
|
-40 to -125 mmHg or -.77 to -2.4 PSI -5.3 to -16.6 kPa
|
|
Therapy Delivery Modes:
|
Continuous & Intermittent
|
|
Battery Type:
|
Alkaline
|
|
Battery Life:
|
3 weeks or more
|
|
Patient Enclosure
|
|
|
Leakage Current:
|
N/A (Battery powered device)
|
|
Storage Conditions:
|
Temperature Range: -20 °C to +60 °C
|
|
Relative Humidity Range: 0 to 95% non-condensing
|
|
Operating Conditions:
|
Temperature Range: +5 °C to +40 °C
|
|
Relative Humidity Range: 0 to 95% non-condensing
|
|
Altitude Range
|
Operating: -1000 ft to +10,000 ft
|
|
Storage:-1000 ft to +18,000 ft
|
|
IEC Classification
|
- Medical Equipment - Equipment not suitable for use in the presence of flammable anesthetic mixture - Type B, Applied Part - Class II - IPX0
|
CAUTION: Federal (USA) law restricts these devices to sale by or on the order of a physician
ELECTROMAGNETIC COMPATIBILITY
Electromagnetic Compatibility- Medical electrical equipment needs special precautions regarding electromagnetic compatibility (EMC) and needs to be installed and put into service according to the EMC information provided in this user manual. Portable and mobile radio frequency (RF) communications equipment can affect medical electrical equipment.
|
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
|
|
The NPD 1000 is intended for use in the electromagnetic environments specified below. The customer or the user of the NPD 1000 should assure it is used in such an environment.
|
|
Emissions Test
|
Compliance
|
Guidance
|
|
RF Emissions CISPR 11
|
Group 1
|
The NPD 1000 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
|
|
RF Emissions CISPR 11
|
Class B
|
The NPD 1000 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
|
|
Harmonic Emissions IEC 61000-3-2
|
Not Applicable
|
|
|
Voltage Fluctuations/ Flicker Emissions IEC 61000-3-3
|
Not Applicable
|
|
|
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
|
|
The NPD 1000 is intended for use in the electromagnetic environments specified below. The customer or the user of the NPD 1000 should assure it is used in such an environment.
|
|
Immunity Test
|
IEC 60601 Test Level
|
Compliance Level
|
Guidance
|
|
Electrostatic Discharge IEC 61000- 4-2
|
± 6 kV Contact ± 8 kV Air
|
± 6 kV Contact ± 8 kV Air
|
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%
|
|
Electrical Fast Transient/ Burst IEC 61000-4-4
|
Not Applicable
|
Not Applicable
|
The NPD 1000 has no connection to Mains
|
|
Surge IEC 61000-4-5
|
Not Applicable
|
Not Applicable
|
The NPD 1000 has no connection to Mains
|
|
Voltage Dips, Short Interrupts, & Variations on Power Supply Lines IEC 61000-4-11
|
Not Applicable
|
Not Applicable
|
The NPD 1000 has no connection to Mains
|
|
Power Frequency Magnetic Fields IEC 61000-4-8
|
3 A/m
|
3 A/m
|
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment
|
ELECTROMAGNETIC COMPATIBILITY CON’T
|
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
|
|
The NPD 1000 is intended for use in the electromagnetic environments specified below. The customer or the user of the NPD 1000 should assure it is used in such an environment.
|
|
Immunity Test
|
IEC 60601 Test Level
|
Compliance Level
|
Guidance
|
|